TS-DW Thermo Shaker
Thermo shaker for shaking and incubating deep well plates.
Allows operating it as 3 independent devices:
- Incubator
- Plate Shaker
- Thermo Shaker
TS-DW provides excellent temperature uniformity across the plate due to patented two-sided heating of the block and the lid, contour heating of the block and close proximity of heating elements to plate walls.
Request a quoteThermo Shaker for Deep Well Plates
Application fields
- Сytochemistry - for in situ reactions
- Immunochemistry - for immunofermentative reactions
- Biochemistry - for enzyme and protein analysis
- Molecular biology - for nucleic acid extraction
Benefits
- Soft or intensive sample shaking
- Rotation speed regulation, stabilization and indication
- Even rotation amplitude throughout the Thermo Shaker platform
- Exceptional temperature uniformity across the plate
- Required operation time setting and indication
- Automatic stopping of the platform movement after expiration of the set time
- Setting and indication of the required temperature on the platform
- A variety of changeable blocks that can accommodate most popular deepwell plates
- Automatic fault diagnostics (temperature sensor, platform heating, lid heating etc.)
Separate blocks to accommodate different Deep Well plates
Available Blocks
- Eppendorf® 96/1000 μl, Eppendorf 96/500 μl,
- Axygen® 96/2200 μl, Axygen® 96/600 μl,
- Riplate® 96/1000 μl,
- Sarstedt® Megablock 96/2200 μl,
- AbgeneTM Storage Plate 96/800 μl,
- AbgeneTM 96/1200 μl,
- Porvair® 96/2000 μl,
- Nunc® 96/2000 μl,
- Riplate® 96/2000 μl,
- Axygen® 96/2200 μl,
- PlateOne® 96/2000 μl,
- Starlab® 96/1200 μl,
- KingFisher™ Deepwell 96 Plate,
- Slicprepᵀᴹ 96,
- PowerMag® Glass Bead Plate 96,
- cluster SSI Bio,
- Vector-Best® 96/1000 μl.
Order No. TS-DW | 105.2300 |
Temperature setting range | +25°C ... +100°C |
Temperature control range | 5°C above ambient ... +100°C |
Temperature setting resolution | 0.1°C |
Temperature uniformity at +37°C | ±0.1°C* |
Temperature accuracy at +37°C | ±0.5°C* |
Time of platform heating from +25°С to +37°С | 6 min* |
* for B-2E thermoblock | - |
Temperature calibration coefficient range | 0.936 ... 1.063 (± 0.063) |
Speed control range | 250 – 1400 rpm |
Orbit | 2 mm |
Display | LCD, 16 × 2 signs |
Digital time setting | 1 min – 96 hrs (1 min increment) |
Timer sound signal | Yes |
Overall dimensions (W×D×H) | 240 x 260 x 160 mm |
Weight | 5.1 kg |
Input current/power consumption | 12 V, 4.8 A / 58 W |
External power supply | Input AC 100 - 240 V 50/60 Hz; Output DC 12 V |
*Temperature Calibration Function
With the help of the temperature calibration function the user can calibrate the unit approx. ±6% of the selected temperature to compensate differences in the thermal behaviour of plates from different manufacturers.
Document