Multi Channel Bioreactor RTS-8 PLUS

RTS-8 PLUS uses the patented ReverseSpin® technology. This is an innovative and energy-saving mixing technique that leads to highly efficient mixing and oxygen supply for aerobic cultivation. By rotating the Falcontubes around their own axis and changing the direction of rotation, the cell suspension is optimally mixed.

Combined with a near-infrared, fluorescence and luminescence measurement systems, it is possible to register cell growth kinetics, pH and O2 non-invasively in real time. For pH and O2, innovative single-use sensor spots are used inside the tubes.

Typical Applications
  • Real-Time cell growth logging
  • Real-Time pH and O2 measurement and logging
  • 3D graphical representation of OD or growth rate over time over unit
  • Pause option
  • Save/Load option
  • Report option: PDF and Excel
  • Remote monitoring option (requires internet connection)
  • Cycling/Profiling options
  • User manual calibration possibility for most cells
Request a quote

Multi Channel Bioreactor RTS-8 PLUS

Although O2 supply is one of the major issues in the cultivation of aerobic organisms, especially in oxygen limited conditions, adequate methods for real monitoring of dissolved oxygen were missing, and sufficient O2 supply was usually assumed.
Innovative non-invasive oxygen sensors integrated in falcon tubes now enable online oxygen monitoring and give new insights into metabolic activities.

The pH is one of the major issues in the cultivation of cells, yeast or bacteria. Cultivation vessels which are sensor limited are widely applied in academic and industrial bioprocess development. As adequate methods for real monitoring of pH were not available, cumbersome at-line sampling was used lacking high data density and interfering with growth.
Non-invasive real time pH measurement provides new insights into metabolic activity and changes in metabolic pathways.

To fully use RTS-8 PLUS capabilities, the device must be connected to a PC and RTS-8 PLUS software. The device cannot be used as a standalone unit.

Specifications 

  • The parallel cultivation of eight tubes makes it possible to save time and resources for bioprocess optimization.
  • Individually controlled bioreactor accelerates optimization process
    Possibility to cultivate microaerophilic and obligate anaerobic microorganisms (not strict anaerobic conditions)
  • Reverse-Spin® mixing principle enables non-invasive biomass measurement in real time
  • Near-infrared optical system makes it possible to register cell growth kinetics
  • Free of charge software for storage, demonstration and analysis of data in real time
  • Compact design with low profile and small footprint for personal application
  • Individual temperature control for bioprocess applications
  • Active cooling for rapid temperature control, e.g. for temperature fluctuation experiments
  • Task profiling for process automatization
  • Cloud data storage to remotely monitor the process of cultivation while at home or using
    a mobile phone
  • Non-invasive O2 and pH measurement allows for accurate monitoring of metabolic activities

Software Features RTS-8 PLUS

  • Real-Time cell growth logging
  • 3D graphical representation of OD or growth rate over time over unit
  • Pause option
  • Save/Load option
  • Report option: PDF and Excel
  • Remote monitoring option (requires internet connection)
  • Cycling/Profiling options
  • User manual calibration possibility for most cells

Advantages of the Sensor Spots

  • They are small
  • Their signal does not depend on the flow rate of the sample
  • They can be physically divided from the measuring system which allows a non-invasive measurement
  • They can be used in disposables
  • Therefore, they are ideally suited for the examination of small sample volumes, for highly parallelized measurements in disposables, and for biotechnological applications

General Technical Data RTS-8 PLUS

Light sourceLaser
Measurement wavelength (λ)850 ± 15 nm
Measurement range0 – 100 OD600
E.coli factory calibration measurement range0 – 50 OD600
S.cerevisiae factory calibration measurement range0  – 75 OD600
Achievable user calibration measurement error (range 0.1-6 OD600)± 0.3
Achievable user calibration measurement error
(range 6-50 OD600)
≤ 5%
Achievable user calibration measurement error
(range 50-75 OD600)
≤ 10%
Measurement periodicity per hour1 – 60
Temperature setting range+15°C bis +60°C
Temperature control range15°C  below ambient up to +60°C
Temperature stability±0.3°C
Sample temperature accuracy (20°C – 37°C)±1°C
Tube sockets8
Sample working volume range3 – 50 ml
Speed control range150 – 2700 rpm
Reverse spin time setting range 150-250 rpm0 sec
Reverse spin time setting range 250–300 rpm2 – 60 sec
Reverse spin time setting range 300–2700 rpm0 – 60 sec
DisplayLCD
Minimum PC requirementsIntel/AMD Processor, 1 GB RAM, Windows Vista/7/8/8.1/10/11, 
USB 2.0 port
Dimensions (W x D x H)350 x 690 x 300 mm
Weight20 kg
Internal power supplyAC 230 V 50 Hz
Power consumption3.15 A / 500 W

O2 Sensor

Range0 – 100%
Accuracy± 0.01% O2 at 0.21%,
± 0.1% O2 at 20.9%
Drift< 0.015% O2 per day
Temperature rangeup to 40°C

pH Sensor

RangepH 4.0 – pH 8.5
Accuracy0,05 pH at pH 7 with onepoint calibration,
± 0,10 pH at pH 7 with precalibration
Drift< 0.005 pH per Day
Temperature range pHup to 40°C