Shaker-Incubator Orbital ES-20
The ES-20 is a compact bench-top shaker incubator used for mixing of biological liquids as well as for incubation and cultivation of biological liquids according to the operator set program.
Built-in microprocessor thermocontroller provides constant temperature control in the incubator chamber.
Request a quoteES-20 Orbital Shaker + Incubator
ES 20 with built–in microprocessor thermocontroller provides constant temperature control in the incubator chamber. Forced heated air circulation inside the transparent plexiglas chamber guarantees even temperature distribution. Dismountable construction makes transportation easy.
Orbital shaking is controlled by the digital tachometer (rpm) and Digital time setting regardless of the temperature. The unit is equipped with the direct–drive system ensuring most reliable stable long–time operation (up to 30 day nights).
The ES-20 is extremely easy to operate, with very straightforward setup of temperature, speed and time, using the two line set–up and status display, which clearly indicates both set and actual values for each of the three parameters.
Five interchangeable platforms for:
(Platforms see below under accessories)
- Growing cell cultures in flasks and other laboratory glassware;
- Extracting tissue samples at physiological temperatures;
- Other sample preparation processes.
Order No. ES-20 | 104.3500 |
Temperature setting range | +25°C ... + 42°C |
Temperature control range | 5°C above ambient ... + 42°C |
Setting resolution | 0.1°C / 10 rpm |
Temperature stability | ±0.5°C |
Temperature accuracy at +37°C | ±0.5°C |
Temperature uniformity at +37°C | ±0.5°C |
Speed control range | 50 – 250 rpm |
Orbit | 10 mm |
Display | LCD, 2 x 16 signs |
Digital time setting | 1 min–96 hrs / non-stop |
Timer sound signal | Yes |
Plexiglas walls thickness | 7 mm |
Load capacity | 2.5 kg |
Overall dimensions (W×D×H) | 340 x 340 x 435 mm |
Internal chamber dimensions | 305 x 260 x 250 mm |
Weight | 13.2 kg |
Nominal operating voltage | 230 V, 50/60 Hz |
Power consumption | 160 W (0.7 A) |
Ist es möglich, das Gerät zur Medienerwärmung in Petrischalen zu verwenden?
Ja, die Petrischalen müssen jedoch abgedeckt werden, um ein Austrocknen zu verhindern.
Ist es möglich, zusätzliche Halterungen für die UP-12-Plattform zu bestellen?
Ja, bitte bestellen Sie HB-200 (s. Technische Daten). Die zusätzliche Haltestange für UP-12 hat die Artikelnummer 105.3513.
Beinhaltet die maximale Belastung das Gewicht von Plattform und Matte?
Nein, die angegebene maximale Belastung gilt nur für die Nutzlast. Sie berücksichtigt nicht das Gewicht von Plattform und Matte.