Thermo Shaker TS-100C / Smart
Thermo Shaker TS-100C provides intensive mixing and temperature control of samples in microtest tubes or PCR plates. This model of thermoshaker differs from TS-100 with a possibility of cooling samples down to +4°C.
The instrument is applicable in:
- Genetic analysis — in extraction of DNA, RNA and further sample preparation;
- Biochemical study of enzymatic reactions and processes;
- Extraction of metabolites from cellular material.
With six interchangeable blocks (see Technical Data).
New: Model TS-100C Smart with Bluetooth connection.
The exact article number can be found in the technical data.
Request a quoteThermo Shaker TS-100C
- Fast reaching of specified mixing speed and maintenance of equal amplitude of rotation throughout the thermoshaker block;
- Stability of maintaining the preset temperature in a wide range throughout the Thermoshaker’s block surface;
- LCD display indicates preset and current values of temperature, speed and time of operation;
- Quiet motor operation, compact size, prolonged service life.
- Temperature Calibration Function
With the help of the temperature calibration function the user can calibrate the unit approx. ±6% of the selected temperature to compensate differences in the thermal behaviour of tubes from different manufacturers.
Functions of heating and mixing can be performed both simultaneously and independently. There are six heating and cooling blocks available, including a block with a plastic lid for the PCR plates. All blocks are mutually interchangeable and can be easily installed on thermoshaker.
There are five heating and cooling blocks available, including a block with a plastic lid for the PCR-plates. All blocks are mutually interchangeable and can be easily installed on Thermo–Shaker. (see Technical Data).
Two Models
- TS-100C: (Order number: 104.1800)
Operation via touch pad - TS-100C Smart: (Order number: 104.1805)
PC-controlled with Bluetooth connection or operation via touch pad
New: Model TS-100C Smart
Software features
- Rotation speed
- Temperature
- Time
- Sound signal
- Creating Profiling programs using controlled parameters
- Visualization of temperature vs time and speed vs time graphs
- Data export to Excel and CSV formats
Error messages / Fault diagnostics
Order Number TS-100C Blocks must be ordered separately! (see below) | 104.1800 (Basic unit without block) |
Order Number TS-100-C Smart Blocks must be ordered separately! (see below) | 104.1805 (Basic unit without block) |
Temperature setting range | +4°C up to +100°C |
Temperature control range | 15°C below ambient ... +100°C |
Temperature setting resolution | 0.1°C |
Temperature stability | ±0.1°C |
Temperature accuracy at +37°C | ±0.5°C |
Temperature uniformity over the block | -> |
at +4°C | ±0.6°C |
at +37°C | ±0.1°C |
at +100°C | ±0.3°C |
Average heating speed from +25°C to +100°C | 5°C/min |
Average cooling speed from +100°C to +25°C | 5°C/min |
Average cooling speed from +25°C to +4°C | 1.8°C/min |
Temperature calibration coefficient range | 0.936...1.063 (± 0.063) |
Speed control range | 250 – 1400 rpm |
Orbit | 2 mm |
Display | LCD, 16 x 2 signs |
Microprocessor controlled temperature, mixing speed and operation time | Yes |
Digital time setting | 1 min – 96 hrs (1 min increment) |
Maximum continuous operation time | max. 168 hours |
Overall dimensions (W×D×H) | 220 x 240 x 90 mm |
Weight | 3.7 kg |
Input current/power consumption | 12 V, 4.9 A / 60 W |
External power supply | Input AC 100 - 240 V; 50/60 Hz; Output DC 12 V |
Block Specifications
Block | Capacity | Order No. | |
---|---|---|---|
SC-18C | 20 x 0.5 ml + 12 x 1.5 ml microtubes | 104.1857 | |
SC-18/02C | 20 x 0.2 ml microtubes + 12 x 1.5ml microtubes | 104.1858 | |
SC-24C | 24 x 2 ml microtubes | 104.1801 | |
SC-24NC | 24 x 1.5 ml microtubes | 104.1802 | |
SC-96AC for PCR- Plates | 96-well unskirted or semi-skirted microplate (0.2 ml) | 104.1803 |
Are there any blocks for Deep well plates?
No, there are no blocks for deep well plates for this instrument. TS-100 is designed to provide high temperature uniformity, stability and accuracy in microtubes and PCR plates. Please, see TS-DW, Thermo-Shaker for deep well plates.
Is any block included in the standard supply set?
No, we offer 5 blocks you can choose from to select best suited for your application. Blocks should be ordered separately. Choose blocks at Technical Data.
It takes around 15 minutes.
Yes, it does. Calibration procedure is described in the user manual point 6.